Resistance to Crown Displacement On A
Hexagonal Implant Abutment
Norman Kwan, DDS, MS, Yang, Silvia: The
Canadian Dental Implant Institute, 206 King St., St. Catharines,
ON L2R 3J7, Canada; Didier Guillaume, DMD; Hoda Aboyoussef, DMD,
MS; Scott D. Ganz, DMD; Saul Weiner, DDS, Department of Restorative
Dentistry, New Jersey Dental School
110 Bergen Street, University Heights, Newark, NJ, 07103, USA
Abstract:
The purpose of this study was to evaluate the resistance and retention
of a hexagonal abutment to crown displacement with varying crown
heights (10, 12, and 14 mm) and axis of load.
Implants were embedded in resin blocks at a 30° angle to the
vertical. Crowns were made at heights of 10, 12, and 14 mm. Groups
of 5 crowns for each of the 3 crown heights were loaded (200 N)
both under the long axis and off axis. The marginal gaps were measured
using standardized periapical radiographs before and after loading.
The gaps were statistically compared using a 3-way analysis of variance
with factors of load, crown height, and point of load. Crowns were
loaded to 10^6 cycles or point of failure.
The 10-mm crowns did not show any displacement. The 12-mm and
14-mm groups only showed displacement of the off axis-loaded crowns.
The average marginal displacement was 193.56 µ (standard deviation
[SD] ± 138.62 µ ) at point of failure (320,717 cycles)
and 400.18 µ (SD ± 644.31) at point of failure (134,278
cycles), respectively.
A 1-piece implant with a standardized abutment designcan provide
sufficient resistance and retention form for crowns of varying dimension
if loads are centered over the long axis of the crown. (Implant
Dent 2004;13:112-119)