  
A
Dynamic Functional Comparative Analysis of Screw Retained vs. Cement
Retained Crowns under Cyclic Fatigue Testing
Norman Kwan, DDS, MS*; Shi Bin, DDS, MS**
Abstract:
A comparative study was conducted to investigate
the stability of identical crowns retained by cement and screw under
dynamic cyclic fatigue conditions. Forty identical crowns were each
telescopically inserted onto hexagonal abutments connected to CP
titanium threaded implants via an abutment screw torqued to 25 Ncm.
Twenty of the crowns were cemented onto the abutment using HY Bond
Zinc Phosphate Cement (Shofu Inc., Japan) while the other twenty
crowns were secured to the abutment implant assembly with a retaining
screw (1.4 x 4mm). Each assembly was embedded in an acrylic block
that fits into the cell holder. The cell holder oscillates at 24?
from the vertical at 0.02 Hz by a custom designed mechanism that
simulates the chewing function. A calibrated Instron Material Testing
Machine (Instron, MS, USA) exerted a variable load set at 15 kg
at 14Hz on the crown during the test.
Crown mobility was periodically assessed in
PTV values using the Periotest (Siemens AG, Bensheim, Germany).
The endpoint was defined when the PTV >10, or when the sample
reached 20M cycles.
Preliminary results indicate that the initial
PTV of the cement and screw retained samples were not significantly
different using the t-test with P>0.05. There is also no significant
difference in the stability of the two groups. The cement retained
group failed on average at 2.60M cycles (SD=2.27M cycles) for n=18,
while screw retained samples failed at 2.17M cycles (SD=1.27M cycles)
when n=20 (P>0.05). Two cement samples were omitted as a result
of machine error. The mode of failure for the two groups is striking.
Of the eighteen cemented samples, fifteen fractured at either the
fixture or abutment screw level. Cemented samples were more rigid
throughout the test and would fracture unexpectedly. In screw retained
samples, the retaining screw would gradually become lose and were
detected by increasing PTV values. There is also a consistent decrease
in the residual torque of the abutment screw. When the screw samples
were left to run continuously, it increased the probability of either
fixture or abutment screw fracture. These results indicate that
the load concentration in the two groups is different under the
same test conditions.
1. Weinberg LA. The Biomechanics of force
distribution in implant-supported prostheses. Int J Oral Maxillofac
Implants, 1993; 8: 19.
2. Preiskel HM, Tsolka P. Telescopic prostheses for implants. Int
J Oral Maxillofac Implants 1998; 13:352-357.
* Director
Canadian Dental Implant Institute
206 King Street, St. Catharines, Ontario
Canada |
** Assistant
Professor
Dept. of Prosthodontics
Stomatological College & Hospital
Hubei Medical University, China |
|